Abstract

The performance of the asymmetrical flow field-flow fractionation channel has been improved by the use of a much thinner (0.12-mm) channel than before and by flow programming (stepwise gradient elution). The thinner channel contributes to a decreased zone broadening which enables complete resolution in a shorter time. Three protein peaks, representing molecular weights from 12 000 to 136 000, were completely resolved within 3 min. Flow programming speeds up the elution of the high-molecular-weight materials which occur late in a fractogram. This enabled separation of a small plasmid fragment (700 base pairs) from the large amounts of a big fragment (4600 base pairs) in 30 min and improved detection of a presumed trimer of albumin. Two viruses (1.8 · 10 6 and 50 · 10 6 daltons) were eluted as narrow peaks within 5 min.