Abstract

Post-column chemical environment modification can affect detection sensitivity and signal appearance when capillary electrophoresis is coupled through electrospray ionization to mass spectrometry (CE-ESI-MS). In this study, changes in the signal intensity and peak shape of N-Acetylneuraminic acid (Neu5Ac) were examined when the modifier solution used in a flow-through microvial interface for CE-ESI-MS was prepared using an acidic or basic background electrolyte (BGE) composition. The use of a basic modifier resulted in improved detection compared to the results obtained when an acidic modifier was used in negative ion mode. Increased sensitivity and more symmetrical peak shape were obtained. Using an acidic modifier, the LOD of Neu5Ac was 47.7nM, whereas for a basic modifier, the LOD of Neu5Ac was 5.20nM. The calculated asymmetry factor at 100nM of Neu5Ac ranged from 0.71 to 1.5 when an acidic modifier was used, while the factor ranged from 1.0 to 1.1 when a basic modifier was used. Properly chosen post-column chemical modification can have a significant effect on the performance of the CE-MS system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.