Abstract

Bioearosol mass spectrometry (BAMS) analyzes single particles in real time from ambient air, placing strict demands on instrument sensitivity. Modeling of the BAMS reflectron time of flight (TOF) with SIMION revealed design limitations associated with ion transmission and instrument sensitivity at higher masses. Design and implementation of a BAMS linear TOF with electrostatic ion guide and delayed extraction capabilities has greatly increased the sensitivity and mass range relative to the reflectron design. Initial experimental assessment of the new instrument design revealed improved sensitivity at high masses as illustrated when using standard particles of cytochrome C (m/z approximately 12,000), from which the compound's monomer, dimer (m/z approximately 24,000) and trimer (m/z approximately 36,000) were readily detected.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.