Abstract
We describe a simple implementation of a slit scanning confocal microscope to obtain an axial resolution better than that of a point-scanning confocal microscope. Under slit illumination, images of a fluorescent object are captured using an array detector instead of a line detector so that out-of-focus light is recorded and then subtracted from the adjacent images. Axial resolution after background subtraction is 2.2 times better than the slit confocal resolution, and out-of-focus image suppression is calculated to attenuate with defocus faster by 1 order of magnitude than in the point confocal case.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.