Abstract

Roentgen stereophotogrammetric analysis (RSA) has been used for over 25 years for accurate micromotion measurement in a wide variety of orthopaedic applications. This study investigated two possible improvements to the method. First, direct linear transformation (DLT) was compared to the traditional RSA reconstruction algorithm. The two methods were considered with respect to standard extrapolation and interpolation calibration cages. Matlab simulations showed that reconstruction accuracy was greatly improved (>60%) by combining DLT with an even distribution of enclosing calibration markers. Second, a benchtop study using phantoms translated at 0.0254-mm intervals showed initial-calibration, followed by removal of the interpolation cage for subsequent exposures, was potentially twice as accurate as self-calibration with an extrapolation cage. These results showed optimizations for the application of RSA when unobstructed space is required.