Abstract

We compared the evaluation accuracy of amplitude envelope statistics under the transmission and reception conditions of compounded plane wave imaging (CPWI) and focused beam imaging (FBI). In a basic study using a homogeneous phantom, we found that the amplitude gradient in the depth direction and the point spread function in the lateral direction spread in the FBI reduced the accuracy of evaluation in amplitude envelope statistics. On the other hand, CPWI showed a more stable evaluation than FBI because of the elimination of sound field characteristics. In CPWI, the multi-Rayleigh model discriminated signals from two types of scatterer with high accuracy in the evaluation using phantoms mimicking fatty liver. It was confirmed that the combination of CPWI and the multi-Rayleigh model is effective for detecting early fatty liver disease. The results show that CPWI is effective for improving the robustness of amplitude envelope statistics.