Abstract
AbstractThe electrical sensing zone technique is long established, well researched and widely used. It is often used as a reference method against which others are compared or calibrated. This position should not be allowed to disguise the potential which exists for still greater accuracy and versatility. In this paper the use of the hydrodynamic focussing technique is evaluated and an experimental comparison is made with a normal Coulter stand for several samples. It is clearly shown that hydrodynamic focussing enables higher resolution and accuracy to be obtained for materials with narrow or unusual particle size distributions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.