Abstract

UV/ozone and wet chemical oxides are compared in terms of reproducibility for scanning capacitance microscopy (SCM). The epitaxial staircase test structure consisted of n-type doped layers with concentrations ranging from 5×10 14 to 5×10 19 cm −3 grown on a uniformly doped Si substrate. Either metal- or diamond-coated probes were used to compare their respective performance. The influence of the tip-coating layer on the reproducibility of measurements with regard to the oxide quality is discussed. In a single scan, the signal variation is mainly owing to the oxide quality and the UV/ozone oxidation method is found to be of higher quality than wet chemical oxidation. In contrast, in multiple scans, the signal variation is strongly dependent on the tip wear and it is clearly evidenced that the diamond-coated probe is suitable for its stability and reliability. The combination of the UV/ozone oxidation process with the use of a diamond tip is an efficient compromise for improved reproducibility of SCM measurements on silicon.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.