Abstract

The study compared the clinical reliability of using a bipolar epicardial wire (6495, Medtronic) over a unipolar type (FEP15, Ethicon) for post-operative pacing in coronary artery surgery. Atrial and ventricular wires of both types were implanted in 18 patients. Sensitivities and pacing thresholds were tested for 5 consecutive days. Results show that pacing thresholds were better maintained with the bipolar wire in both atria and ventricles. However, sensing failures were frequent in the atrial position (34% vs 9.3% compared with unipolar). By contrast, in the ventricle, no sensing failures occurred (0% vs 17.6% compared with unipolar). Furthermore, sensing magnitude was significantly better (11.13±1.32 vs 5.65±0.53 mV, P<0.001). We conclude that a single 6495 bipolar wire is effective for temporary ventricular pacing, whilst double unipolar wires remain a useful strategy for securing atrial sensing and pacing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.