Abstract

This study investigates the use of multivariate linear regression to estimate three bipolar ECG leads from the 12-lead ECG in order to improve P-wave signal strength. The study population consisted of body surface potential maps recorded from 229 healthy subjects. P-waves were then isolated and population based transformation weights developed. A derived P-lead (measured between the right sternoclavicular joint and midway along the costal margin in line with the seventh intercostal space) demonstrated significant improvement in median P-wave root mean square (RMS) signal strength when compared to lead II (94μV vs. 76μV, p<0.001). A derived ES lead (from the EASI lead system) also showed small but significant improvement in median P-wave RMS (79μV vs. 76μV, p=0.0054). Finally, a derived modified Lewis lead did not improve median P-wave RMS when compared to lead II. However, this derived lead improved atrioventricular RMS ratio. P-wave leads derived from the 12-lead ECG can improve signal-to-noise ratio of the P-wave; this may improve the performance of detection algorithms that rely on P-wave analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.