Abstract

The presence of monospecific dense fine speckled 70 (DFS70) pattern in indirect immunofluorescence assay (IFA) without concomitant systemic autoimmune related diseases (SARD)-associated antibodies could be an exclusion biomarker for SARD. Since interpretation of DFS pattern on IFA is subjective, an assay for confirming the presence of anti-DFS70 is required. We evaluated the performance of two commercial assays [fluorescence enzyme immunoassay (FEIA) and line immunoassay (LIA)] for detecting anti-DFS70. Sera with monospecific DFS (n=176) and without DFS (n=179) pattern from referred patients for ANA testing, in two independent laboratories and healthy donors, were investigated for anti-DFS70 by FEIA (Phadia EliA) and LIA (Euroimmun). The assay performance including sensitivity and specificity at different cut-offs was evaluated, and optimal cut-offs were determined. The newly established optimal cut-offs (2.7 U/mL on EliA, band intensity of 28 on LIA) showed significantly better assay performance in detecting anti-DFS70 and confirming monospecific DFS pattern. A relative sensitivity of 93.7% with relative specificity of 100% on EliA and a relative sensitivity of 96.6% with relative specificity of 95.3% on LIA were achieved. Superior Cohen's Kappa agreements with IFA were also achieved for both assays (0.936 for EliA and 0.906 for LIA). Application of an optimal cut-off can significantly increase the assay performance for anti-DFS70 and enhance the accuracy in reporting DFS pattern by IFA.