Abstract
The erratic performance of the plasma jet in the analysis of magnesium in blood plasma was found to arise from the accumulation of residue in the atomization annulus. This problem was circumvented by the use of an inexpensive Zeiss atomizer which was easily dismantled and cleaned. The plasma jet, equipped with this atomizer, was successfully applied to the routine analysis of magnesium in blood plasma. Compared to both medium- and large-bore Beckman atomizers the Zeiss atomizer provided similar sensitivity and greatly improved repeatability. The plasma jet, equipped with a Zeiss atomizer, may be universally applicable to the analysis of many solutions, particularly those containing salts in high concentration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Applied Spectroscopy
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.