Abstract

We propose a technique for myocardial motion estimation based on image registration using both B-mode echocardiographic images and tissue Doppler sequences acquired interleaved. The velocity field is modeled continuously using B-splines and the spatiotemporal transform is constrained to be diffeomorphic. Images before scan conversion are used to improve the accuracy of the estimation. The similarity measure includes a model of the speckle pattern distribution of B-mode images. It also penalizes the disagreement between tissue Doppler velocities and the estimated velocity field. Registration accuracy is evaluated and compared to other alternatives using a realistic synthetic dataset, obtaining mean displacement errors of about 1 mm. Finally, the method is demonstrated on data acquired from six volunteers, both at rest and during exercise. Robustness is tested against low image quality and fast heart rates during exercise. Results show that our method provides a robust motion estimate in these situations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.