Abstract

To improve the computer determination of thoracic gas volume (TGV), two new approaches were worked out. 1) A new program was designed, which overcomes the difficulties encountered in the time recognition of the panting maneuver and rules out the artifacts. Such a procedure is based on the data analysis in the pressure-volume time derivatives plane. 2) A hyperbolic fitting of the signals recorded during the panting maneuver was introduced. This last procedure, lying on Boyle's law, has proved to be useful in case of large mouth pressure changes. In fact the error induced by the conventional linear fitting may reach 500 ml (9% of the TGV value).