Abstract

Measurement of malondialdehyde (MDA) is an important contribution to the assessment of oxidative stress. We report a sensitive and reproducible high-performance liquid chromatography (HPLC) method for measurement of plasma MDA in the assessment of lipid peroxidation. Using methyl malondialdehyde (Me-MDA) as an internal standard with reversed-phase HPLC and UV detection and derivatisation with 2,4 dinitrophenylhydrazine (DNPH), we obtained maximum MDA values with 60-min incubation of 10% plasma with 1 M NaOH at 60 °C. The dilution of the plasma and a longer incubation time in the alkaline hydrolysis step greatly improved recovery of MDA from its bound form. Ratios of peak height of MDA/Me-MDA were linear over a range of 0–100 μ M with correlation coefficients >0.99. The recovery was 88.5%. Within and between run variations were <4 and <7%, respectively. The mean MDA value measured in 20 healthy volunteers was 13.8 μ M (±1.32).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call