Abstract
Abstract Arrhythmic patients treated with quinidine have been found to ingest caffeine from various sources. Caffeine interferes with the quantitation of quinidine in plasma samples. This method for determination of quinidine in plasma is based on high performance isocratic liquid chromatography with the use of a C-18 bonded reverse phase column at room temperature. Unlike some liquid chromatographic procedures for quinidine, caffeine does not interfere. Quinidine is extracted from alkalinized plasma into benzene. The benzene layer is removed and evaporated to dryness under nitrogen at 50 C. The residue is reconstituted with methanol and injected into the column. The mobile phase is 30/70 (v/v) mixture of methanol/0.4% glacial acetic acid. Analysis can be completed within 10 minutes. The procedure is sensitive (0.5mg/L) and is well reproducible (CV= 3.5% for a 2 mg/L concentration in plasma).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.