Abstract

Stereotactic radiosurgery techniques for a linear accelerator typically use circular radiation fields to produce an essentially spherical radiation distribution with a steep dose gradient. Target volumes are frequently irregular in shape, and circular distributions may irradiate normal tissues to high dose as well as the target volume. Improvements to the dose distribution have been made using multiple target points and optimizing the dose per arc to the target. A retrospective review of 20 radiosurgery patients has suggested that the use of elliptically shaped fields may further improve the match of the radiation distribution to the intended target volume. This hypothesis has been verified with film measurements of the radiation distribution obtained using elliptical radiation beam in a head phantom. Reductions of 40% of the high dose volume have been obtained with elliptical fields compared to circular fields without compromising the dose to the target volume.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.