Abstract

The aim of the study was to investigate the feasibility of low-tube-voltage computed tomography (CT) in combination with 3-dimensional adaptive iterative dose reduction (AIDR-3D) algorithm in lower extremity CT angiography. The CT data of the 120-kV group (n = 30) were reconstructed with filtered back projection (FBP) algorithm. The CT data of the 100-kV group (n = 30) were reconstructed with FBP as well as AIDR-3D algorithms. The 100-kV group showed significantly lower dose-length product than the 120-kV group (P < 0.05). In comparison with the 120-kV and FBP protocol, the 100-kV and FBP protocol showed significantly increased vascular density and noise (P < 0.05). However, in the 100-kV group, images reconstructed with AIDR-3D showed significantly lower noise and significantly higher signal-to-noise ratio and contrast-to-noise ratio than FBP (P < 0.05). Low-tube-voltage (100 kV) 320-row CT in combination with AIDR-3D reconstruction can significantly improve the image quality and reduce radiation dose of lower extremity CT angiography.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.