Abstract

A highly sensitive and selective HPLC method was developed for the determination of amoxicillin in human plasma. After addition of buffer and internal standard, the sample was ultrafiltered and injected on to a precolumn to remove polar plasma interferences. Detection was effected with a UV detector set at 230 nm. The limit of quantification for amoxicillin was 50.1 ng/ml with an imprecision of 4.2% using 0.25 ml of plasma. Linearity was confirmed over the whole calibration range (25.4-0.0501 μg/ml) and the inter-day variation ranged from 2.0 to 4.5%. The method was validated according to GLP guidelines and its suitability was demonstrated by the analysis of several hundred samples in a bioequivalence study. The method can be used to determine pharmacokinetic parameters of amoxicillin in humans after a single oral dose of 500 mg.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.