Abstract

In order to eliminate the use of liquid helium for the extraction of atmospheric gases from polar ice cores, two units of a redesigned top load helium exchange gas cryostat were built and tested. The cryostats feature the shortest and largest diameter sample wells built to date, a base temperature below 7 Kelvin, and a sample well without baffles. The cryostats allowed shortening the length and thus increasing the gas pressure inside our sample tubes by 58% and increasing the amount of sample ending up in the mass spectrometer by 4.4%. The cryostats can either be used as mobile stand‐alone units for manual gas processing lines or integrated into a fully automated vacuum extraction and gas analysis line. For the latter application the cryostat was equipped with a custom‐designed automated changeover system.