Abstract

Four compact planetary ultraviolet spectrographs have been built by Southwest Research Institute and successfully operated on different planetary missions. These spectrographs underwent a series of ground radiometric calibrations before delivery to their respective spacecraft. In three of the four cases, the in-flight measured sensitivity was approximately 50% lower than the ground measurement. Recent tests in the Southwest Research Institute Ultraviolet Radiometric Calibration Facility (UV-RCF) explain the discrepancy between ground and flight results. Revised ground calibration results are presented for the Rosetta-Alice, New Horizons-Alice, the Lunar Reconnaissance Orbiter Lyman- Alpha Mapping Project, and Juno-Ultraviolet Spectrograph (UVS) and are then compared to the original ground and flight calibrations. The improved understanding of the calibration system reported here will result in improved ground calibration of the upcoming Jupiter Icy Moons Explorer (JUICE)-UVS.