Abstract

Abstract Heart failure (HF) is a chronic disease with high hospitalizations rates and increased mortality. The six-minute walk test (6MWT) is a standardized method to assess the functional health of HF patients by measuring the distance walked in six minutes [1]. The improved 6MWT with new filter algorithms, based on the development of smartphone application [9], can be used everywhere with a good global navigation satellite system (GNSS) signal. This paper tests multiple filtering approaches to reduce the measured distance error in the application.We were able to decrease the average relative error of the application from 4.5% to 2.0%. Additionally, the number of experiments with relative error over our chosen clinically significant difference of 5% is reduced by 90%. Keeping these results in mind, the system can be considered clinically acceptable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.