Abstract

The gas-liquid chromatographic method previously reported by us was improved with regard to the gas chromatographic columns and their operating conditions. It became possible to perform more accurate quantification of the 22 protein amino acids, including asparagine and glutamine, in less than 25 min on a dual set of columns. At the same time, the method allowed determination of the amino acids in the range 0.5-10 μg of each by eliminating the contaminant peaks derived from solvents and reaction vials. A single sample could be analysed in an overall time of 1 h.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.