Abstract

To demonstrate improved frequency selective fat suppression in MRI using a magnetic susceptibility matching foam by reducing B0 inhomogeneities induced within the body by air-tissue interfaces. Flexible pyrolytic graphite (PG) composite foam was tailored to match the magnetic susceptibility of human tissue and was shaped to surround the cervical spine region. Field maps and frequency selective fat suppressed T1 -weighted FLASH images were acquired at 3 Tesla in both phantoms and six healthy necks. B0 field uniformity was shimmed to a target critical threshold of 1 ppm for fat suppression. The percentage of voxels in the phantom that did not achieve the critical threshold was reduced from 64% without the PG foam to only 1% with the foam. A similar decrease from 16% to 2% was observed in the in vivo region of interest. PG foam improved B0 field uniformity by moving air-tissue field gradients outside of the neck where they cannot cause MRI artifacts. The PG foams consistently mitigated signal dropout, improved overall SNR, and enabled more robust frequency selective fat suppression.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.