Abstract

We inserted, in 8 dogs, implants with either porous-coated or grit-blasted titanium surface and coated with hydroxyapatite (HA) into trabecular bone in the proximal humerus, using a 1 mm gap model. After 25 weeks, push-out tests showed that energy absorption for porous-coated implants was twice that of grit-blasted implants, whereas shear stiffness was reduced by one fifth, indicating a stronger fixation of porous-coated implants. Macroscopically, all grit-blasted implants had delamination of the HA coating, whereas porous implants failed mostly at the HA-tissue interface. Porous-coated implants had 47% bone ingrowth and grit-blasted implants 70% (p = 0.02), however, no difference in absolute surface area was found. Part of the HA coating was resorbed during the implantation period as regards volume and thickness. HA coverage was more reduced on porous-coated than on grit-blasted implants (p = 0.01). No foreign-body reaction or osteolysis was seen. An important finding was that one fifth of the surface with complete resorption of HA coating was replaced by newly formed bone.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.