Abstract

A new entrance optic for a Brewer spectrophotometer has been designed and tested both in the laboratory and during solar measurements. The integrated cosine response deviates by 2.4% from the ideal, with an uncertainty of +/- 1%. The systematic uncertainties of global solar irradiance measurements with this new entrance optic are considerably reduced compared with measurements with the traditional design. Simultaneous solar irradiance measurements between the Brewer spectrophotometer and a spectroradiometer equipped with a state-of-the-art shaped diffuser agreed to within +/- 2% during a five-day measurement period.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.