Abstract

Deep brain stimulation (DBS) devices have been implanted for treatment of basic tremor, Parkinson's disease and dystonia. These devices use electrodes in contact with tissue to deliver electrical pulses to targeted cells, to elicit specific therapeutic responses. In general, the neuromodulation industry has been evolving towards smaller, less invasive electrodes. However, current electrode materials do not support small sizes without severely restricting the stimulus output. Hence, an improved electrode material will benefit present and future DBS systems. In this study, five DBS leads were modified using a cost-effective and materials-efficient process for applying an ultra-low impedance platinum-iridium alloy coating. One DBS lead was used for insertion test and four DBS leads were chronically pulsed for 12 weeks. The platinum-iridium alloy significantly improved the electrical properties of the DBS electrodes and was robust to insertion into brain and to 12 weeks of chronic pulsing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.