A graphite-paste tyrosinase biosensor was improved by adding 1-methoxyphenazine methosulfate as a mediator. Mediator modification enhanced sensitivity to phenol 4-fold and long-term stability 3-fold. Phenol could be detected at 25 nM (S/N = 2) using an Ag/AgCl reference electrode. The biosensor was used to measure the activity of a toxicologically significant enzyme, neuropathy target esterase (NTE), which yields phenol by hydrolysis of the substrate, phenyl valerate. Using the new biosensor, blood and brain NTE inhibition by organophosphorus (OP) compounds with different neuropathic potencies were well correlated (r = 0.990, n = 7), supporting the use of blood NTE as a biochemical marker of exposure to neuropathic OP compounds.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.