Abstract

To improve the accuracy of ventricular volume estimates and effective temporal resolution in fast cardiac acquisitions. Data sets at intermediate temporal phases were generated by means of sharing of views from two temporally adjacent data sets. A simulated model was used, and studies with patients and volunteers were conducted. View sharing was implemented in both fast gradient-echo and fast phase-contrast cine acquisitions; breath holding was used when possible. Uniform repetition time (TR) radio-frequency (RF) excitation allowed a better assessment of the end-diastolic ventricular volume. In addition, view sharing provided a better estimate of end-systolic ventricular volume in cases in which rapid changes in volume occurred at or about the temporal boundary of the source images. View sharing also provided a much smoother representation of dynamic cardiac motion when viewed in a cine loop. View sharing and uniform TR RF excitation improve the accuracy of end-systolic and end-diastolic ventricular volume measurements by improving the effective temporal resolution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.