Abstract

To evaluate the value of adding single-shot balanced turbo field-echo (b-TFE) sequence to conventional magnetic resonance cholangiopancreatography (MRCP) for the detection of common bile duct (CBD) stone. One hundred thirty-seven consecutive patients with suspected CBD stone underwent MRCP including single-shot b-TFE sequence. Twenty-five patients were confirmed with CBD stone by endoscopic retrograde cholangiopancreatography or ultrasonography. Two radiologists reviewed two image protocols: protocol A (conventional MRCP protocol: unenhanced T1-, T2-, and respiratory-triggered three-dimensional fat-suppressed single-shot turbo spin-echoMRCP sequence) and protocol B (protocol A plus single-shot b-TFE sequence). The sensitivity, specificity, positive (PPV) and negative predictive value (NPV), and area under the receiver-operating-characteristic (ROC) curve (AUC) for the detection of CBD stone were compared. The sensitivity (72%) and NPV (94%) were the same between the two protocols. However, protocol B was greater in the specificity (99%) and PPV (94%) than protocol A (92% and 67%, respectively) (P=0.0078 and 0.031, respectively). The AUC was significantly greater for protocol B (0.93) than for protocol A (0.86) (P=0.026). Inclusion of single-shot b-TFE sequence to conventional MRCP significantly improved the specificity and PPV for the detection of CBD stone.