Abstract

We compared 2 continuous flow resectoscopes for transurethral resection of the prostate—a new intrinsic bladder pressure-controlled drainage system and the suction drainage system. In 31 patients undergoing transurethral prostatic resection who were assigned randomly to a pressure-controlled (17) or suction drainage (14) resectoscope group the required irrigant volume and operative blood loss (milliliter per gram resected tissue) were appreciably less with pressure-controlled drainage, while the resection rate (gram per minute) was significantly greater. Irrigant absorption, as evidenced by weight gain and changes in hematocrit, serum osmolality and serum sodium, was minimal in either group, with no significant difference between the 2 methods. In addition, there was no significant difference in intraoperative intravesical pressure between the 2 groups. The suction drainage system led to pump drainage failure in certain rotational positions, resulting in diminished visibility and increased resection time. The pressure-controlled drainage system required no suction devices or adjustments, thus, allowing constant improved visibility and uninterrupted resection.These results suggest that the efficiency and ease of transurethral prostatic resection are superior with the pressure-controlled continuous flow drainage system than with the suction drainage system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.