Abstract

A new cleanup method for large lipid samples is described for the analysis of organochlorine traces. The method allows both electron capture gas chromatographic (GC) quantitation of sub-ppb levels and confirmation by GC-mass spectrometry. The lipid is extracted with concentrated sulfuric acid from a 2-5% solution of light petroleum ether. The method has been scaled up without increasing the petroleum volume by the alternate addition of lipids and sulfuric acid. Recovery studies were performed with 11 common organochlorine contaminants spiked in butterfat and in vegetable oils at the ppb level. The method is characterized by low lipid carryover of 0.02% by minimal use of solvent (about 2-3 mL/g lipid), and by its simplicity. A 50-200 g sample can be cleaned up in one day. Five micrograms of incurred p,p'-DDE has been cleaned up from a 50 g butterfat sample and purified by one TLC run. The eluted compound was analyzed by GC-MS and showed one peak (flame photometric detector) with a mass spectrum identical to authentic p,p'-DDE.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.