Abstract

The calibration of our interference microscope is currently performed by an elaborate multi step process. As a result the total uncertainty of a measurement performed with the interference microscope is much larger than the intrinsic repeatability of the microscope which is of the order of 1 nm. A major contribution to the total uncertainty is a length dependent factor, resulting from a calibration step using gauge blocks that finally yields 8 nm uncertainty for a step height of 2 micrometers . In order to reduce the total uncertainty we propose a novel step-height standard and calibration procedure. The standard is adjustable and can be simultaneously measured with the interference microscope and a laser interferometer allowing calibration of the entire dynamic range of the microscope with a single artefact. The new calibration method eliminates the process that contributes most to the total uncertainty budget in the current procedure. A possible implementation of the step-height standard is presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.