Abstract

To evaluate the impact of deep learning (DL) reconstruction in enhancing image quality and nerve conspicuity in LSP MRN using DESS sequences. Additionally, a geometric image combination (GIC) method to improve DESS signals' combination was proposed. Adult patients undergoing 3.0 Tesla LSP MRN with DESS were prospectively enrolled. The 3D DESS echoes were separately reconstructed with and without DL and DL-GIC combined reconstructions. In a subset of patients, 3D T2-weighted short tau inversion recovery (STIR-T2w) sequences were also acquired. Three radiologists rated 4 image stacks ('DESS S2', 'DESS S2 DL', 'DESS GIC DL' and 'STIR-T2w DL') for bulk motion, vascular suppression, nerve fascicular architecture, and overall nerve conspicuity. Relative SNR, nerve-to-muscle, -fat, and -vessel contrast ratios were measured. Statistical analysis included ANOVA and Wilcoxon signed-rank tests. p < 0.05 was considered statistically significant. Forty patients (22 females; mean age = 48.6 ± 18.5years) were enrolled. Quantitatively, 'DESS GIC DL' demonstrated superior relative SNR (p < 0.001), while 'DESS S2 DL' exhibited superior nerve-to-background contrast ratio (p value range: 0.002 to < 0.001). Qualitatively, DESS provided superior vascular suppression and depiction of sciatic nerve fascicular architecture but more bulk motion as compared to 'STIR-T2w DL'. 'DESS GIC DL' demonstrated better nerve visualization for several smaller, distal nerve segments than 'DESS S2 DL' and 'STIR-T2w DL'. Application of a DL reconstruction with geometric image combination in DESS MRN improves nerve conspicuity of the LSP, especially for its smaller branch nerves.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call