Abstract

We investigated the characteristics of the array detector system and TLD for dose distribution by the moving wedge in therapeutic photon beams. Also we have improved an electrometer for the array detector system for radiation therapy. In dynamic field irradiation, the moving wedge technique and its fraction methods are available through the computer-controlled asymmetric independent collimator. The array detector system consists of an electrometer, a solid detector, TLD and an array phantom. It is used for the point dose measurement and the field size scanning. We evaluated the dosimetric characteristics of the virtual wedge and the conventional fixed wedge by the scanning system. In our analysis for the 6 and 10 MV photon beams, the maximum dose in the virtual-wedge and the fixed-wedge factors varied from 1.2 % to 1.6 % for square collimator setting ranging from 10 to 20 cm, respectively.KeywordsArray detectorWedge fieldRadiation dosimetry

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.