Abstract

To assess the influence of frame rate settings on longitudinal strain (LS) and mechanical synchrony (SYN) values in Speckle Tracking Echocardiography (STE) of healthy fetuses. In this prospective study, we collected transversal or apical four-chamber-views of 121 healthy fetuses between 20 and 38weeks of gestation using three different frame rate (FR) settings (≥ 110, 100±10, 60±10 frames per second). We assessed the segmental and the global LS of both ventricles (2C) and of the left ventricle (LV) offline with QLab 10.8 (Philips Medical Systems, Andover, MA, USA). Inter- and intraventricular SYN were calculated as time difference in peak myocardial strain between the mid-segments of left and right ventricle (interventricular, 2C_Syn) and lateral wall and septum of the left ventricle (intraventricular, LV_Syn), respectively. In 84.3% STE was feasible at all three FR settings. The LS increased in both views at higher FRs to a statistically noticeable extent. SYN measurements and the absolute differences at patient level between the FR settings showed no statistically noticeable alterations. STE is feasible at low and high FR settings. SYN emerges to be a robust parameter for fetal STE as it is less affected by the FR. High FRs enable high temporal resolutions and thus an accurate examination of fetal hearts. Future research for the technical implementation of tailored fetal STE software is necessary for reliable clinical application.