Abstract
When helical tomotherapy plans are transferred between two matched machines, the leaf opening times are changed. A published method of checking these is based on the differences in leaf latency between the two machines. We have extended this method to account also for the differences in leaf fluence opening factors (LFOF) between the two machines. When LFOF is not taken into account, the number of false negatives (with under 95% of opening times agreeing within 5 ms or under 99% agreeing within 10 ms) is approximately 10% of patients. After correcting for LFOFs, all transfers meet the criteria.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.