Abstract

The implementation of a laser pump/X-ray probe scheme for performing picosecond-resolution X-ray diffraction at the 1W2B wiggler beamline at Beijing Synchrotron Radiation Facility is reported. With the hybrid fill pattern in top-up mode, a pixel array X-ray detector was optimized to gate out the signal from the singlet bunch with interval 85 ns from the bunch train. The singlet pulse intensity is ∼2.5 × 10(6) photons pulse(-1) at 10 keV. The laser pulse is synchronized to this singlet bunch at a 1 kHz repetition rate. A polycapillary X-ray lens was used for secondary focusing to obtain a 72 µm (FWHM) X-ray spot. Transient photo-induced strain in BiFeO3 film was observed at a ∼150 ps time resolution for demonstration.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call