Abstract
True continuous axial bed motion has been implemented on a high-resolution positron emission tomography (PET) scanner for use in both two-dimensional (2-D) and three-dimensional (3-D) acquisition modes. This has been accomplished by modifications in the bed motion controller firmware and by acquiring data in list mode. The new bed controller firmware was shown to provide an accurate and constant bed speed down to 0.25 mm/s with a moderate patient weight load. The constant bed motion eliminates previously reported dead-time due to bed positioning when using small discrete bed steps. The continuous bed motion was tested on uniform phantoms, in 2-D and 3-D. As a result of the continuous axial motion, a uniform axial sensitivity is achieved. This was also reflected in the reconstructed images, which showed an improvement in axial image uniformity (1.4% for continuous sampling, 5.0% for discrete) as well as an improvement in %SD uniformity in comparison to conventional step-and-shoot acquisitions. The use of the continuous axial motion also provide slight improvements in 2-D emission and transmission scanning, resulting in an overall improved image quality in whole-body PET.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.