Abstract

BackgroundCardiopulmonary resuscitation (CPR) metrics including compression rate and depth are associated with improved outcomes and the need for high-quality CPR is emphasized in both the American Heart Association (AHA) and Heart and Stroke Foundation of Canada (HSFC) guidelines. While these metrics can be utilized to assess the quality of CPR, they are infrequently measured in an objective fashion in the emergency department.ObjectivesAs part of an Emergency Department (ED) Quality Improvement (QI) project, we sought to determine the impact of real-time audio-visual (AV) feedback during CPR amongst ED healthcare providers.MethodsParticipants performed two minutes of uninterrupted CPR without AV feedback, followed by two minutes of CPR with AV feedback after a two-minute rest period in a simulated CPR setting. CPR metrics were captured by the defibrillator and uploaded to review software for analysis of each event.ResultsThe use of real-time AV feedback resulted in a significant improvement in the number of participants meeting AHA/HSFC recommended depth (38%, p = 0.0003) and rate (35%, p = 0.0002). Importantly, ‘compressions in target’, where participants met both rate and depth simultaneously, improved with AV feedback (19 vs 61%, p < 0.0001).ConclusionsWe found a significant improvement in compliance with CPR depth and rate targets as well as ‘compressions in target’ with the use of real-time AV feedback during simulation training. Future research is needed to ascertain whether these results would be replicated in other settings. Our findings do provide a robust argument for the implementation of real-time AV CPR feedback in Hamilton Emergency Departments.

Highlights

  • Cardiopulmonary resuscitation (CPR) metrics including compression rate and depth are associated with improved outcomes and can be utilized to assess the quality of CPR administered [1,2,3,4]

  • The use of real-time AV feedback resulted in a significant improvement in the number of participants meeting American Heart Association (AHA)/Heart and Stroke Foundation of Canada (HSFC) recommended depth (38%, p = 0.0003) and rate (35%, p = 0.0002)

  • We found a significant improvement in compliance with CPR depth and rate targets as well as ‘compressions in target’ with the use of real-time AV feedback during simulation training

Read more

Summary

Introduction

Cardiopulmonary resuscitation (CPR) metrics including compression rate and depth are associated with improved outcomes and can be utilized to assess the quality of CPR administered [1,2,3,4]. While the American Heart Association (AHA) and Heart & Stroke Foundation of Canada (HSFC) guidelines emphasize the need for high-quality CPR, this is infrequently measured in an objective fashion in the. Cardiopulmonary resuscitation (CPR) metrics including compression rate and depth are associated with improved outcomes and the need for high-quality CPR is emphasized in both the American Heart Association (AHA) and Heart and Stroke Foundation of Canada (HSFC) guidelines. While these metrics can be utilized to assess the quality of CPR, they are infrequently measured in an objective fashion in the emergency department

Objectives
Methods
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.