Abstract

An automatic, miniature and multi-parameter on-line water quality monitoring system based on a micro-spectrometer is designed and implemented. The system is integrated with the flow-batch analysis and spectrophotometric detection method. The effectiveness of the system is tested by measuring chemical oxygen demand (COD) and ammonia-nitrogen in water. The results show that the modified system provides a cost-effective, sensitive, reproducible and reliable way to measure COD and ammonia-nitrogen in water samples with automatic operation and low toxic chemical consumption. In addition, the experiment results show that the relative error of the system is less than 10%, the limit of detection is 2 mg/L COD and 0.032 mg/L ammonia-nitrogen, respectively, and the relative standard deviation was 6.6% at 15.0 mg/L COD (n = 7) and 5.0% at 0.300 mg/L ammonia-nitrogen (n = 7). Results from the newly designed system are consistent with the data collected through the Chinese national standard analysis methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.