Abstract

Background: Automated cerebrospinal fluid (CSF) drainage systems allow for the mobilization of patients with an external CSF drain. The aim of this study is to describe the implementation of an automated CSF drainage system in neurosurgical patients with external CSF drains. Methods: A feasibility study was performed using an automated CSF drainage system (LiquoGuard®7, Möller Medical GmbH, Fulda, Germany) in adult neurosurgical patients treated with external lumbar or external ventricular drains between December 2017 and June 2020. Limited mobilization was allowed—patients were allowed to adjust their inclined beds, sit in chairs and walk under the supervision of a nurse or physical therapist. The primary outcome was the number of prematurely terminated drainage sessions. Results: Twenty-three patients were included. Drainage was terminated prematurely in eight (35%) patients. In three (13%) of these patients, drainage was terminated due to signs of hydrocephalus. Pressure-controlled drainage in patients with external lumbar drains (ELD) showed inaccurate pressure curves, which was solved by using volume-controlled drainage in ELD patients. Conclusion: The implementation of an automated CSF drainage system (LiquoGuard®7) for CSF drainage allows for early mobilization in a subset of patients with external CSF drains. External lumbar drains require volume-based drainage rather than differential pressure-dependent drainage.

Highlights

  • The aim of this study is to describe the implementation of the LiquoGuard®7 in mobilized hydrocephalus patients with external cerebrospinal fluid (CSF) drains

  • Nineteen EVDs and five external lumbar drains (ELD) were connected to the LiquoGuard® 7

  • One drain dislodgement occurred while connected to the LiquoGuard® 7

Read more

Summary

Introduction

Temporary treatment of hydrocephalus using external lumbar or ventricular drainage is common in neurosurgical practice. Conventional hydrostatic pressure-controlled drainage with dripping chambers are the gold standard [1]. Automated cerebrospinal fluid (CSF) drainage systems allow for the mobilization of patients with an external CSF drain. The aim of this study is to describe the implementation of an automated CSF drainage system in neurosurgical patients with external CSF drains. Methods: A feasibility study was performed using an automated CSF drainage system (LiquoGuard® 7, Möller Medical GmbH, Fulda, Germany) in adult neurosurgical patients treated with external lumbar or external ventricular drains between December 2017 and June 2020. Drainage was terminated prematurely in eight (35%) patients. In three (13%) of these patients, drainage was terminated due to signs of hydrocephalus. Pressure-controlled drainage in patients with external lumbar drains (ELD)

Objectives
Results
Discussion
Conclusion