Abstract

Charge detection mass spectrometry (CDMS) depends on the measurement of the charge induced on a cylinder by individual ions by means of a charge-sensitive amplifier. Electrical noise limits the accuracy of the charge measurement and the smallest charge that can be detected. Thermal noise in the feedback resistor is a major source of electrical noise. We describe the implementation of a charge-sensitive amplifier without a feedback resistor. The design has significantly reduced 1/f noise facilitating the detection of high m/z ions and substantially reducing the measurement time required to achieve almost perfect charge accuracy. With the new design we have been able to detect individual ions carrying a single charge. This is an important milestone in the development of CDMS.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.