Abstract

An in vitro radiostereometric analysis (RSA) phantom study of a total knee replacement was carried out to evaluate the effect of implementing two new modifications to the conventional RSA procedure: (i) adding a landmark of the tibial component as an implant marker and (ii) defining an implant-based coordinate system constructed from implant landmarks for the calculation of migration results. The motivation for these two modifications were (i) to improve the representation of the implant by the markers by including the stem tip marker which increases the marker distribution (ii) to recover clinical RSA study cases with insufficient numbers of markers visible in the implant polyethylene and (iii) to eliminate errors in migration calculations due to misalignment of the anatomical axes with the RSA global coordinate system. The translational and rotational phantom studies showed no loss of accuracy with the two new measurement methods. The RSA system employing these methods has a precision of better than 0.05mm for translations and 0.03° for rotations, and an accuracy of 0.05mm for translations and 0.15° for rotations. These results indicate that the new methods to improve the interpretability, relevance, and standardization of the results do not compromise precision and accuracy, and are suitable for application to clinical data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.