Abstract

Golay Complementary Sequence (GCS) is conventionally regarded as not applicable to blood flow detection due to potential side-lobe motion artifact in the decoding. In our previous study, it has been shown that GCS decoding can be performed by low-pass filtering in slow-time Doppler domain with cutoff frequency of PRF/4 to alleviate the motion artifact. In this work, the proposed GCS Doppler decoding has been implemented in a pre-clinical high-frequency ultrasound imaging system to study the color Doppler quality and signal-to-noise ratio (SNR) using phantom and in-vivo experiments. The phantom results show that, when the Doppler shift is within ±PRF/4, the GCS provides comparable Doppler information to conventional un-coded method while the SNR improves. The in-vivo results also demonstrate that Doppler penetration increases from 13.5 mm to 14.2 mm in the mouse's kidney. For the abdominal aorta, GCS Doppler also enables the increase of transmit frequency from 30 MHz to 35 MHz to detect blood flow.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.