Abstract

FOCI pulses are a variant of hyperbolic secant pulses in which the RF amplitude, RF frequency, and gradient waveform are all modulated by the same function A(t). This increases the usable gradient amplitude without requiring a corresponding increase in RF amplitude. In this paper the implementation and inversion profiles of FOCI pulses on a clinical MR system are described, showing improved slice definition and chemical-shift offset behavior. Their adiabatic behavior is confirmed, and their use with an ISIS sequence for localized MRS is illustrated. Finally the effects of waveform digitization are considered, and the implications for SAR and dB/dt are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.