Abstract

The use of stent grafts for endovascular aortic repair has become an important treatment option for aortic aneurysms requiring surgery. This treatment has achieved excellent outcomes; however, problems like type 1 endoleaks and stent graft migration remain. Bio stent grafts (BSGs), which are self-expanding stents covered with connective tissue, were previously developed using "in-body tissue architecture" technology. We assessed their early adaptation to the aorta after transcatheter implantation in a beagle model. BSGs were prepared by subcutaneous embedding of acryl rods mounted with self-expanding nitinol stents in three beagles for 4weeks (n=3/dog). The BSGs were implanted as allografts into infrarenal abdominal aortas via the femoral artery of three other beagles. After 1month of implantation, aortography revealed no stenosis or aneurysmal changes. The luminal surface of the BSGs was completely covered with neointimal tissue, including endothelialization, without any thrombus formation. The cover tissue could fuse the luminal surface of the native aorta with tight conjunctions even at both ends of the stents, resulting in complete impregnation of the strut into the reconstructed vascular wall, which is expected to prevent endoleaks and migration in clinical applications.