Abstract

How dynamic activity in neural circuits gives rise to behavior is a major area of interest in neuroscience. A key experimental approach for addressing this question involves measuring extracellular neuronal activity in awake, behaving animals. Recently developed Neuropixels probes have provided a step change in recording neural activity in large tissue volumes with high spatiotemporal resolution. This protocol describes the chronic implantation of Neuropixels probes in mice and rats using compact and reusable 3D-printed fixtures. The fixtures facilitate stable chronic in vivo recordings in freely behaving rats and mice. They consist of two parts: a covered main body and a skull connector. Single-, dual- and movable-probe fixture variants are available. After completing an experiment, probes are safely recovered for reimplantation by a dedicated retrieval mechanism. Fixture assembly and surgical implantation typically take 4-5 h, and probe retrieval takes ~30 min, followed by 12 h of incubation in probe cleaning agent. The duration of data acquisition depends on the type of behavioral experiment. Since our protocol enables stable, chronic recordings over weeks, it enables longitudinal large-scale single-unit data to be routinely obtained in a cost-efficient manner, which will facilitate many studies in systems neuroscience.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.