Abstract

Establish the surgical procedure for subretinal implantation of multiple photovoltaic arrays for the restoration of sight. Multiple silicon photovoltaic arrays of 1 mm in diameter and 30 µm in thickness were implanted subretinally via single retinotomy in rabbits. Ophthalmoscopic imaging and optical coherence tomography (OCT) were used to validate the implants' placement. Vitrectomy, followed by subretinal fluid injection for retinal detachment and retinotomy, allowed accurate placement of seven modules in the bleb, covering approximately a 3.5-mm diameter area on the retina via a single 1.5-mm retinotomy. OCT confirmed complete reattachment of the retina over the implants. Subretinal implantation of multiple photovoltaic arrays via a single retinotomy, followed by their tiling, minimizes the scleral and retinal incisions and provides better fit to the spherical shape of the eye ball, compared to a single, larger module. Such minimally traumatic procedure can be performed with 20-gauge intraocular instruments.