Abstract

Purpose: To describe a contralateral femoral approach for iliac branch device implantation using a steerable sheath in the setting of an existing bifurcated stent-graft. Technique: The method is demonstrated in an 80-year-old man who developed a 4-cm iliac aneurysm 3 years after implantation of an Endurant bifurcated stent-graft. Both femoral arteries were cannulated after surgical cutdown. The steerable sheath was advanced from the contralateral side over the neobifurcation of the bifurcated stent-graft. A 0.014-inch Roadrunner wire was used as a through-and-through wire to stabilize the curve of the sheath and to get proper push. The bridging stent-graft for the iliac branch was advanced over this sheath to seal the iliac aneurysm. During the entire procedure, the sheath was stable over the neobifurcation without pulling it down. Conclusion: The contralateral femoral approach for iliac branch graft implantation is feasible in cases with an extant bifurcated stent-graft using a steerable sheath and a through-and-through wire.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.